Flexibly adjustable mounts

For the installation of the BLE Radar , flexibly adjustable mounts are available for various installation situations (see Assembly sets). The alignment of the BLE Radar with the flume to be measured is carried out for all flexibly adjustable mounts using the procedure described below.

Note: Avoid touching the spherical joint and ensure that the ball of the spherical joint is also not contaminated otherwise with grease, oil or lubricants of any kind. A such contamination reduces the friction between the ball and the swivel arm, which results in the fact that a retained stable alignment can no longer be guaranteed. In case of contamination, the ball can be cleaned using isopropyl or other degreasing cleaning agents.

Aligning the BLE Radar

1 Spirit level
2 BLE Radar
3 Sensor recess
4 Laser meter Bosch GLM50-27CG (301757)
5 Laser meter adapter Bosch PRO GLM 50-xxx (301756, optional accessory)
  1. In the first instance, install the sensor recess suitable for your installation situation. An explanation of this is provided in the chapters Wall mounting (flexibly adjustable mount), Pipe mounting, Strut/rod mounting und Installation in a manhole cover DN600.
  2. Roughly align the swivel arm of the sensor recess with the flume to be determined. If necessary, unlock the corresponding clamping screw.
  3. Insert the BLE Radar in the sensor recess of the mount and fasten it using the G 1½" union nut included in the scope of supply of the BLE Radar .

    Inserting the BLE Radar in the sensor recess

    1 BLE Radar
    2 Sensor recess
    3 G 1½" union nut
  4. The following step is only required if you want to use a laser distance meter when aligning the BLE Radar .
  5. Slide the optional accessory "Laser meter adapter Bosch PRO GLM 50-xxx " (301756) on the intended bracket of the sensor recess, then insert the already activated laser distance meter in the adapter.

     

    The following laser distance meters are compatible with the adapter:
    • Bosch PRO GLM 50-27 CG
    • Bosch PRO GLM 50-27 C
    • Bosch PRO GLM 50-25 G
    • Bosch PRO GLM 50-23 G
    • Bosch PRO GLM 50-22
    • Bosch PRO GLM 40-31

    How-To-Video: Mounting the laser distance meter

    Mounting the laser distance meter

    1Laser meter Bosch GLM50-27CG (301757)
    2Laser meter adapter Bosch PRO GLM 50-xxx " (301756)
    3Bracket of the sensor recess
  6. Loosen the clamping screw to unlock the spherical joint. Subsequently align the BLE Radar by means of the aid of the laser beam of the distance meter with the spot of the flume where the water level should be determined and use the spirit level of the BLE Radar to align the BLE Radar exactly perpendicular.
    Note: Without the optional laser distance meter, the spirit level of the BLE Radar is the only available mounting aid for the correct alignment of the BLE Radar .

    How-To-Video: Aligning the radar sensor using a laser distance meter

    Aligning the BLE Radar

    1 Clamping screw to unlock the spherical joint
    2 Spirit level
    3 Laser meter Bosch GLM50-27CG (301757)
  7. Fasten the spehrical joint again by tightening the clamping screw.
  8. The following steps are only necessary if a very precise alignment of the BLE Radar is required.
  9. Loosen the G 1½" union nut and remove the BLE Radar from the sensor recess. Then, instead of the BLE Radar , install the "Laser meter adapter Bosch PRO GLM 50-xxx " with the laser distance meter inserted in the sensor recess and check whether the laser beam of the distance meter points to the spot of the flume where you would like to determine the water level.
    Note: In doing so, ensure that the alignment (spehrical joint and swivel arm) of the sensor recess is not altered, provided no readjustment is needed.

    Carry out an alignment correction if the laser beam does not aim for the desired spot. However, in this case you also have to recheck the perpendicular alignment in the next step.

    How-To-Video: Practical use of the laser distance meter

    Inserting the laser distance meter in the sensor recess

    1 Laser meter Bosch GLM50-27CG (301757)
    2 Laser meter adapter Bosch PRO GLM 50-xxx " (301756)
    3 Bracket of the sensor recess
  10. Replace the "Laser meter adapter Bosch PRO GLM 50-xxx " with inserted laser distance meter again with the BLE Radar .
    Note: In doing so, ensure that the alignment (spehrical joint and swivel arm) of the sensor recess is not altered, provided no readjustment is needed.

    If the alignment of the sensor recess has been altered in the previous step, the BLE Radar has to be aligned exactly perpendicular again using the spirit level. Then repeat the previous step and this step until the alignment of the BLE Radar meets your requirements.