Output channels

Configuration section to configure the settings for the 2 relays.

Configuration section "Output channels"

1 freely selectable designation for the output channels
2 Basic setting for the output channel
off Output channel deactivated
digital Digital output

Mode "digital"
3 Inverts the level issued on the device
4 Setpoint (on/off) that should be issued
Note: If one of the check boxes "D1" or "D2" is selected in the "Trigger" tab of the configuration section "Measurement channels" or "Internal channels", the "setpoint" entered is ignored for the corresponding output channel.

Tip:

Additional explanation on "Digital" mode

Invert Setpoint   Output on the device
Off Off = Off (low)
Off On = On (high)
On Off = On (high)
On On = Off (low)

Tip:

Example explaining how to determine the switching state at the output via trigger

Basic settings Measurement cycle 1 min.
Measurement channels -> Trigger   D1 Selected
  Threshold Greater or equal

100

  Hyst % 5
Output channels   Mode digital
  Invert off
  Setpoint on
Trigger not active green line  
Trigger active red line  
Output on the device Switching state  

Initial situation: One of the measurement channels is to be used to determine the switching state at the ouput channel depending on the measurement value using the triggers. If one of the checkboxes "D1" (output channel "RELAY1") or "D2" (output channel "RELAY2") in the "Trigger" tab of the "Measurement channels" configuration section (see Measurement channels ) is selected for one of the measurement channels, the output channel is set to "off (Low)" at the time of the next measurement as long as the trigger for this measurement channel is not active.

Explanation: The output on the device is set to "off (Low)" up to the time of 2, as the measurement value on the measurement channel is below the trigger level. The trigger is active from the time of 2 which sets the output to "on (High)". At the time of 3, the output is ultimately set to "off (Low)" again as the measurement value falls below the "Level - hyst" value (100-5=95).

Ergänzung: If one of the two checkboxes ("D1" or "D2") has been selected, the "setpoint" entered for the relevant output channel via the interface is ignored. The parameter "invert" is however taken into account. If "invert" is active for the output channel, the switching states at the output are exactly the opposite of the example described above (i.e. with trigger active, the output channel is set to "on (High)").