Troubleshooting and repair

General problems

Problem Cause/solution
Communication problems (between PRE-PT-D30 and data logger)
  • Reduce the distance between the PRE-PT-D30 and the data logger, and try to avoid obstacles such as walls and similar.
  • The capacity of the battery is virtually depleted.
The measurement value "Pressure" is not plausible
  • The alignment of the PRE-PT-D30 is incorrect. The process connction must always point upwards.
  • The capacity of the battery is virtually depleted.
  • Air is trapped in the supply line to the PRE-PT-D30 . Vent the supply line. If the PRE-PT-D30 has been installed in accordance with the recommendations in chapter Installation on a piping system fitting, you can use the shut-off valve for this purpose.
  • Check whether the inlet of the process connection is blocked (see Overview). If blockages occur repeatedly, install a sediment separator in the supply line to the PRE-PT-D30 .