Quick start

1/4: Check your order

  1. Please check the order against the delivery slip to ensure it is complete.

    The components illustrated below are required for commissioning and operating the Jellox Node . However, the sets may also contain other components.

    1 MDN Magnet
    2 Jellox Node
    3 Antenna (e.g.Flat antenna Disc Multi Band FME-F 2m )

2/4: Create a site

If you do not have an account yet, you will receive the web address of the server, the access data and any credit balance from your responsible sales partner.
  1. Log in to the server via the web interface.

  2. Click on "Customer". Select an existing customer or create a new one.

    Only if you have an account to manage multiple customers.

    1 Menu item to call up the list of customers 3 List of available customers
    2 Creating a new customer    

  3. Click on "Sites / Applications". Open the wizard for creating a new site by clicking on the plus symbol (2).

    1 Menu item to call up the list of existing sites / applications 2 "Add new site / application" symbol

  4. Follow the wizard instructions through to the "Done" step.

    The serial number required for the Wizard can be found on the type plate of the device.

  5. Review the details in the summary and create the site by clicking the "Create site" button.

    By clicking "create site", the site that is subject to fees will be created.

    1 This amount is debited from the customer's account for this site immediately upon creation and subsequently on a monthly basis.
    2 Checkbox to accept the general terms and conditions
    3 Link to the general terms and conditions
    4 "Create site" button

  6. Review the created site.

    1 Name of the site. Click on the name to open the specific input screen for configuring the site.
    2 Product name as specified on the type plate of the device
    3 Serial number of the device linked to the site
    4 State of the site

3/4: Install remote sensor module

The installation of the remote sensor modules varies depending on the module type.
  1. Install the remote sensor modules of type LVL‑R‑5 (if included)

    For optimum accuracy, install the LVL‑R‑5

    • at a location with calm water surface
    • centrally above the flume
    • in distance from structures and fittings such as ladders or struts
    • perpendicular to the water surface
    • at a min. distance of 30 cm to the water surface.

    A detailed explanation of the installation is provided here

  2. Install the remote sensor modules of type Gauge (if included)

    A detailed explanation of the installation is provided here

  3. Install the remote sensor modules of type VEL‑R‑5 (if included)

    For optimum accuracy, install the VEL‑R‑5

    • at a location with constant flow
    • in the centre of a channel that runs straight for a long distance
    • parallel to the flow
    • against the direction of flow (upstream)
    • at an angle of 45° to the water surface
    • with the "top marking" facing upwards

    A detailed explanation of the installation is provided here

4/4: Commissioning the Jellox Node

  1. Connect the antenna.

    How-To-Video: Connecting the Multi band antenna with bracket

    1Antenna connector2Antenna cable

  2. Click on the name of the site to open the specific input screen for configuring the site.

    1Site name

  3. Start the "Guided commissioning" process by clicking on the corresponing button.

    1Opens the input screen for the guided commissioning process
  4. The "Guided commissioning" wizard first prompts you to establish a connection.

    To do so, hold the MDN Magnet on the spot shown in the following illustration until the symbol on the display that indicates the status of the mobile nework connection starts blinking.

    How-To-Video: Triggering a connection

    1 MDN Magnet 2 Jellox Node
    3 Status of the mobile network connection 4 detected hold of the reed switch

    A detailed explanation of the display is provided here

  5. Open the relevant wizards to configure the individual components one by one and follow the instructions. At last, open the "wizard to complete the set-up" (5).

    Not every component needs to be configured immediately. You can also configure the device later via the individual input windows or you can restart the guided commissioning process.

    1 Opens the wizard for the basic configuration of the measurement system
    2 Opens the wizard to configure a remote sensor module of type LVL-R-5
    3 Opens the wizard to configure a remote sensor module of type Gauge
    4 Opens the wizard to complete the commissioning process

 

Congratulations!

You have successfully commissioned your Jellox Node !

Tell us about your experience with the commissioning process. Your feedback is important to us.

 

Feedback form