Installing the VEL‑R‑5

Important:
  • Ensure installation is completed correctly.
  • Comply with existing legal and/or operational directives.
  • Improper handling can cause injuries and/or damage to the devices.
  • The VEL‑R‑5 is not suitable for outdoor use.
  • When using metal mounting components, these must be integrated into the grounding concept in a suitable manner.

The installation site must be selected according to specific criteria. The following conditions must be avoided in any case:

  • Direct sunlight
  • Direct weather exposure (rain, snow, etc.)
  • Objects that radiate intense heat (maximum ambient temperature: -10...+70 °C )
  • Objects with a strong electromagnetic field (frequency converter or similar)
  • Corrosive chemicals
  • Mechanical impacts
  • Direct installation on paths or roads
  • Vibrations
  • Radioactive emissions