Determining the measurement values
How the determination of the measurement values is triggered and which values are generated depends on whether the BLE Radar has been coupled with a data logger (pairing). The corresponding dependencies are broken down in the following table. Please note that when a measurement command is received, the 4 measurement values "Level", "Distance", "Status of the distance" and "Amplitude" are always generated together. However, the interval at which the data logger sends the measurement commands must not be less than 10 sec. seconds.
| Coupled | Not coupled | |
|---|---|---|
| Water level [mm] | Triggered via measurement command by data logger | 10 sec. interval |
| Distance [mm] | ||
| Status of the distance [ ] | --- | |
| Amplitude[ ] | --- | |
| Temperature of the electronics [0,01 °C] | 60 sec. interval | --- |
| Battery voltage [mV] | --- | |
| Estimated remaining battery life [h] | 24 h interval | --- |