Installing the Jellox Pulse
Important:
- Ensure installation is completed correctly.
- Comply with existing legal and/or operational directives.
- Improper handling can cause injuries and/or damage to the devices.
- The Jellox Pulse is not approved for use in closed drains.
Note: It is recommended that all metal mounting fixtures be
properly earthed.
The installation site must be selected according to specific criteria. The following conditions must be avoided in any case:
- Direct sunlight
- Direct influence of weather (rain, snow, ...)
- Objects that radiate intense heat (maximum ambient temperature: -20...+60 °C )
- Objects with a strong electromagnetic field (frequency converter or similar)
- Corrosive chemicals or gases
- Mechanical impacts
- Direct installation on paths or roads
- Vibrations
- Radioactive emissions
Tip: Approx. 15 cm of space should be left beneath the
device for the cable connections. Further information about the installation dimensions can
be found in the relevant sub-chapter.