Device settings

Basic settings

Input window "Device settings", tab "Basic settings"

1 Buttons for switching between the individual tabs of the input window
2 Selection of the operating mode
normal Operation in accordance with the configuration
transport Transport lock activated. This means that all of the operations (measurement, recording, transmission, etc.) are stopped to minimise energy consumption during transport or storage.
3 Interval the internal measurement values are determined in. (00:00 same as record interval)

This interval is used to monitor the conditions under which a ‘Smart Action’ is triggered that uses one of the internal measurement values.

Tip: If the measurement interval is selected to be longer than the record interval, the measurements are carried out in the record interval.
4 Time between transmissions
5 Time between measurement data recordings
6 Button to apply the settings
Note: To save the settings, the "Save" button below the "Smart Actions" configuration section must also be clicked.

 

Expert settings

Input window "Device settings", tab "Expert settings"

1 Position update interval (00:00 positioning is completed with every connection)
2 Interval in which the sensor modules carry out measurement. (Read Only)

This interval is used to monitor the conditions under which a "Smart Action" is triggered that uses the primary measured value of one of the sensor modules.