Troubleshooting and repair

General problems

Problem Cause/solution
Communication problems (between BLE Gauge and the data logger)
  • Reduce the distance between the BLE Gauge and the data logger, and try to avoid obstacles such as walls and similar.
  • The capacity of the battery is virtually depleted.
  • Check that the correct serial number was specified when linking the BLE Gauge with the data logger (see user manual of the respective data logger).
Fill level is not plausible.
  • Check whether there are any kinks in the cable or whether the bend radius has fallen below the minimum permissible value of 60 mm .
  • Check the pressure compensation membrane of the BLE Gauge for any impurities (see Overview).
  • Check whether the inlet openings of the pressure sensor are blocked (see Overview).
  • Check whether the pressure sensor is immersed in the medium to be measured at least up to the marking (position of the diaphragm) (see Overview).